FDA Adverse Event Malfunction Summary report: N

GREENLIGHT MOXY FIBER OPTIC

MDR report key: 3805525 · Received May 12, 2014

Report

Report Number
2937094-2014-00371
Event Type
Malfunction
Date Received
May 12, 2014
Date of Event
April 3, 2014
Report Date
April 3, 2014
Manufacturer
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
Product Code
GEX
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER (B)(4): THE FIBER SHOWS A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE METAL CAP EXHIBITS VERY MILD DETRITUS ADHESION; THE GLASS CAP EXHIBITS VERY MILD DEVITRIFICATION AT THE OUTPUT WINDOW. BASED ON THE ANALYSIS RESULTS, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH LIMITING THE PERFORMANCE OF THE FIBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROSTATE PROCEDURE, @ 121550 JOULES; 23 MINUTES, THE PHYSICIAN NOTED ¿MIRROR BREAKS.¿ THE PROCEDURE WAS COMPLETED USING A SECOND FIBER. PATIENT OR USER OUTCOME: ¿NO DAMAGES TO THE PATIENT¿ REPORTED. GLAND VOLUME: 45 ML

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
284522 GREENLIGHT MOXY FIBER OPTIC POWERED LASER SURGICAL INSTRUMENT GEX AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) 0010-2400 241A

Patients

Seq Age Sex Outcome Treatment
1