FDA Adverse Event Injury Summary report: N

TECNIS

MDR report key: 3803289 · Received May 9, 2014

Report

Report Number
2648035-2014-00220
Event Type
Injury
Date Received
May 9, 2014
Date of Event
February 27, 2014
Report Date
April 16, 2014
Manufacturer
ABBOTT MEDICAL OPTICS
Product Code
HQL
PMA / PMN Number
P990080
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE EXPLANTED INTRAOCULAR LENS (IOL) WAS RETURNED TO THE MANUFACTURING SITE FOR INVESTIGATION. A VISUAL INSPECTION OF THE RETURNED SAMPLE WAS PERFORMED AT 10X MICROSCOPE MAGNIFICATION. THE VISUAL INSPECTION REVEALED THAT THE LENS WAS RECEIVED WITH THE OPTIC EDGE TORN AND ONE (1) OF LOOPS (HAPTICS) WAS DETACHED. THE LENS FINDINGS OBSERVED COULD BE ASSOCIATED WITH THE HANDLING OF THE LENS DURING THE EXPLANT PROCESS AND HAS BEEN INDICATED AS OPERATIONAL IN THE RESULTS CODE. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED. ALL PROCESS OPERATIONS PRESENTED IN THE MANUFACTURING RECORD FROM GENERATION TO BOXING WERE IN COMPLIANCE. ALL TESTS RESULTS SHOWED A PASS CONDITION. NO QUANTITY DISCREPANCY WAS FOUND. NO DEVIATION OR NON-CONFORMANCE (NCR) WAS GENERATED DURING MANUFACTURING OF THIS PRODUCTION ORDER. THE INSPECTION DATA FOR DIOPTER POWER WAS REVIEWED AND WAS FOUND WITHIN SPECIFICATION. NO ISSUES WERE REPORTED DURING THE MANUFACTURING OF THE LENS, THE LENS MET MANUFACTURING SPECIFICATIONS PRIOR TO RELEASE. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED. PLACEHOLDER.

Additional Manufacturer Narrative · 1

(B)(4). ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED. PLACEHOLDER.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) WAS REQUIRED TO BE EXPLANTED FROM THE LEFT EYE OF A PATIENT IN A SECONDARY PROCEDURE OWING TO DYSPHOTOPSIA. A NEW LENS WAS IMPLANTED IN THE EYE. NO FURTHER INFORMATION HAS BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
280952 TECNIS MONOFOCAL IOLS HQL ABBOTT MEDICAL OPTICS ZA9003

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention