FDA Adverse Event Malfunction Summary report: N

ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM

MDR report key: 3803248 · Received May 9, 2014

Report

Report Number
2024168-2014-02979
Event Type
Malfunction
Date Received
May 9, 2014
Date of Event
March 31, 2014
Report Date
April 15, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
NIO
PMA / PMN Number
P110028
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION AND THE REPORTED PREMATURE DEPLOYMENT WAS CONFIRMED. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES. A QUERY OF THE ELECTRONIC COMPLAINT HANDLING DATABASE REVEALED NO OTHER SIMILAR INCIDENTS REPORTED FROM THIS LOT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION OF THE ABSOLUTE PRO VASCULAR 6.0/40MM/135CM STENT DELIVERY SYSTEM THE STENT WAS OBSERVED TO BE PROTRUDING FROM THE DEPLOYMENT SHEATH. THE DEVICE WAS NOT USED ON THE PATIENT. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281992 ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM SELF EXPANDING STENT SYSTEM NIO AV-TEMECULA-CT 3102561

Patients

Seq Age Sex Outcome Treatment
1