XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 3803212 · Received May 9, 2014

Report

Report Number
2024168-2014-02975
Event Type
Malfunction
Date Received
May 9, 2014
Date of Event
April 14, 2014
Report Date
April 15, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT PRODUCT: GUIDE CATH: GUIDELINER. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE STENT DAMAGE WAS CONFIRMED. THE FAILURE TO ADVANCE AND DIFFICULTY TO REMOVE/GUIDING CATHETER EXTENSION RESISTANCE COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. BASED ON VISUAL, DIMENSIONAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. A QUERY/REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT USING A FEMORAL ARTERY ACCESS APPROACH DURING A PROCEDURE OF THE HEAVILY CALCIFIED, NON-TORTUOUS, 95% STENOSED, DE NOVO, DISTAL LEFT MAIN ARTERY AFTER MULTIPLE PRE-DILATATIONS WITH A 3.0 X 15 MM NON-ABBOTT BALLOON DILATATION CATHETER UP TO 16 BAR, A GUIDELINER WAS USED AND THE 3.0 X 28 MM XIENCE XPEDITION STENT DELIVERY SYSTEM (SDS) WAS ADVANCED BUT COULD NOT CROSS THE LESION. DURING THE ATTEMPT TO REMOVE THE SDS THROUGH THE GUIDELINER RESISTANCE WAS MET; OUTSIDE THE ANATOMY IT WAS NOTED THAT THE STENT STRUTS WERE FLARED AT THE DISTAL END. TWO (2) NON-ABBOTT STENTS WERE SUCCESSFULLY DEPLOYED IN THE PROCEDURE WITHOUT REPORTED PATIENT ISSUE. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281741 XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 3061341

Patients

Seq Age Sex Outcome Treatment
1