FDA Adverse Event Malfunction Summary report: N

HEARTWARE® VENTRICULAR ASSIST SYSTEM

MDR report key: 3803185 · Received May 9, 2014

Report

Report Number
3007042319-2014-00443
Event Type
Malfunction
Date Received
May 9, 2014
Date of Event
August 16, 2013
Report Date
April 9, 2014
Manufacturer
HEARTWARE, INC
Product Code
DSQ
PMA / PMN Number
P100047
Removal / Correction Number
Z-1607-2014
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRELIMINARY EVALUATION AND TESTING REVEALED THAT THE RETURNED BATTERY CONTAINED A FAULTY CELL. THIS FINDING WAS RE-ASSESSED PER OUR SOP REQUIREMENTS, AND DETERMINED TO BE REPORTABLE. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN THIRTY (30) DAYS OF COMPLETION OF THE INVESTIGATION. EVALUATION IN PROGRESS.

Additional Manufacturer Narrative · 1

THE BATTERY WAS RETURNED; VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICE IN RELATION TO THE REPORTED EVENT. THOROUGH EXTERNAL VISUAL INSPECTION OF THE BATTERY REVEALED NO SIGNS OF PHYSICAL DAMAGE OR CONTAMINATION. FUNCTIONAL TESTING REVEALED THAT THE RETURNED BATTERY CONTAINED A FAULTY CELL PAIR, WHICH LIKELY CONTRIBUTED TO THE REPORTED EVENT. AN INTERNAL INVESTIGATION (CAPA) HAS BEEN OPENED TO ADDRESS THE ISSUE. THIS EVENT RELATES TO THE FSCA Z-1607-2014 ISSUED BY HEARTWARE, INC. FOR THIS MATTER.

Description of Event or Problem · 1

APPROXIMATELY FOUR MONTHS POST HVAD IMPLANTATION, THE SITE REPORTED THAT THE PATIENT EXPERIENCED POWER SOURCE CHANGE IN THE CONTROLLER EARLIER THAN EXPECTED. THE BATTERY WAS REMOVED FROM THE PATIENT AND A NEW BATTERY WAS SUPPLIED. NO HARM OR INJURY TO THE PATIENT WAS REPORTED AS A RESULT OF THIS INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281902 HEARTWARE® VENTRICULAR ASSIST SYSTEM CIRCULATORY ASSIST SYSTEM, BATTERY DSQ HEARTWARE, INC

Patients

Seq Age Sex Outcome Treatment
1 (B)(4) - BATTERY| (B)(4) - BATTERY| (B)(4) - BATTERY