FDA Adverse Event Malfunction Summary report: N

STERRAD 100NX STERILIZER

MDR report key: 3803137 · Received May 9, 2014

Report

Report Number
2084725-2014-00181
Event Type
Malfunction
Date Received
May 9, 2014
Date of Event
April 23, 2014
Report Date
April 23, 2014
Manufacturer
ADVANCED STERILIZATION PRODUCTS
Product Code
MLR
PMA / PMN Number
K071385
Removal / Correction Number
Z-1026-1027-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER WAS DISPATCHED TO CUSTOMER SITE. ODOR/SMELL WAS CONFIRMED. A PREVENTATIVE MAINTENANCE (PM2) WAS PERFORMED. UNIT MEETS SPECIFICATIONS AND WAS RETURNED TO SERVICE.

Additional Manufacturer Narrative · 1

ROOT CAUSE IDENTIFIED IN CAPA WAS FOUND TO BE PREMATURE FAILURE/SATURATED OIL MIST FILTER OR VACUUM PUMP OIL ASP INVESTIGATION SUMMARY: THE INVESTIGATION INCLUDED A REVIEW OF THE DEVICE HISTORY RECORD, SERVICE HISTORY, TRENDING OF THE PRODUCT MALFUNCTION CODE, FAILURE MODE AND EFFECTS ANALYSIS, HEALTH HAZARD EVALUATION, SYSTEM HAZARD AND USER MISUSE ANALYSIS, AND CAPA. THE DHR (DEVICE HISTORY RECORD) WAS REVIEWED AND NO ISSUES RELATING THE FAILURE MODE WERE NOTED. THE INVOLVED UNIT MET MANUFACTURER SPECIFICATIONS AT THE TIME OF RELEASE. THE SERVICE HISTORY FOR THIS UNIT FOR THE PAST 6 MONTHS (10/31/2013 TO (B)(6) 2014) DID NOT REVEAL A SIGNIFICANT TREND FOR THIS SAME ISSUE. THE TREND OF THE PRODUCT MALFUNCTION CODE ODOR/SMELLS WAS COMPLETED FROM JUNE 2013 THROUGH (B)(6) 2014 AND REVEALED A SIGNIFICANT TREND WHICH WAS ADDRESSED THROUGH CAPA. THE FMEA (FAILURE MODE AND EFFECTS ANALYSIS) REVEALED THE RISK PRIORITY NUMBER FOR THIS FAILURE MODE IS GREATER THAN THE ACCEPTABLE LIMIT AND WAS ADDRESSED THROUGH CAPA. THE HHE (HEALTH HAZARD EVALUATION) WAS REVIEWED FOR THE RISK OF ODOR AND SMELL EXPOSURE. THE SEVERITY AND OCCURRENCE FOR THE GENERAL POPULATION WERE LIMITED (TRANSIENT, MINOR IMPAIRMENT, NO MEDICAL TREATMENT REQUIRED) AND THE PRODUCT PROBLEM HAS BEEN KNOWN TO RESULT IN THE IDENTIFIED HARM, BUT ONLY OCCASIONALLY AND/OR UNDER UNUSUAL CIRCUMSTANCES. THE SHUMA (SYSTEM HAZARD AND USER MISUSE ANALYSIS) SHOWS THAT THE RISK IS AS LOW AS REASONABLY PRACTICAL FOR EXPOSURE TO ODOR OR ODORANTS. THE CAPA (CORRECTIVE AND PREVENTATIVE ACTION) IDENTIFIED THE ROOT CAUSE FOR THE ODOR/SMELLS ISSUE AS: (1) PREMATURE FAILURE OF THE USED AND SATURATED STERRAD® 100NX OIL MIST FILTER CAUSED OIL VAPOR EMISSIONS THAT EXACERBATED THE ODOR/SMELL COMPLAINTS REPORTED FOR THE STERRAD® 100NX SYSTEM. (2) THE USE OF A LESS OXIDATIVELY STABLE VACUUM PUMP OIL CAUSED THE ODOR/SMELLS FOR THE STERRAD® 100NX SYSTEM. NO PARTS WERE RETURNED FOR FURTHER EVALUATION. ASP WILL CONTINUE TO TRACK AND TREND THIS ISSUE.

Description of Event or Problem · 1

AN INTERNATIONAL CUSTOMER REPORTED AN EVENT OF A "STRONG PEROXIDE ODOR" EMITTING FROM THE STERRAD 100NX STERILIZER. THERE WAS NO REPORT OF ANY INJURIES OR HUMAN REACTIONS. AN ASP FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY EVENT DATED (B)(6) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281663 STERRAD 100NX STERILIZER STERRAD EQUIPMENT (MLR) MLR ADVANCED STERILIZATION PRODUCTS NA 1042070051

Patients

Seq Age Sex Outcome Treatment
1