Narratives
Additional Manufacturer Narrative · 1
USING A WELCH ALLEN BLACK BODY THE IRT-3020 WAS RETESTED. THE BLACK BODY WAS SET TO 36.0 DEGREES CENTIGRADE AND THE IRT-3020 WAS SET TO CAL MODE. THE TEMPERATURE OF THE BLACK BODY WAS TAKEN AND THE READING WAS 55.7 DEGREES CENTIGRADE. THESE ARE FAILING RESULTS. THE THERMOMETER WILL BE SENT TO SUPPLIER FOR ROOT CAUSE ANALYSIS.
Description of Event or Problem · 1
A CONSUMER REPORTED THAT HER THERMOMETER HAD ALLEGEDLY GIVEN AN INACCURATE READING. THE DEVICE WAS READING IN THE NORMAL RANGE DESPITE THEIR SON HAVING A FEVER. THE CHILD WAS GIVEN MOTRIN (ORALLY) AND WAS SEEN BY THEIR PHYSICIAN ON THE FOLLOWING DAY. THERE WERE NO FURTHER COMPLICATIONS AND THE PT IS DOING WELL NOW.