FDA Adverse Event Other Summary report: N

BRAUN

MDR report key: 3802979 · Received May 5, 2014

Report

Report Number
1314800-2014-00035
Event Type
Other
Date Received
May 5, 2014
Date of Event
April 3, 2014
Report Date
May 1, 2014
Manufacturer
KAZ USA, INC.
Product Code
FLL
PMA / PMN Number
K983295
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

USING A WELCH ALLEN BLACK BODY THE IRT-3020 WAS RETESTED. THE BLACK BODY WAS SET TO 36.0 DEGREES CENTIGRADE AND THE IRT-3020 WAS SET TO CAL MODE. THE TEMPERATURE OF THE BLACK BODY WAS TAKEN AND THE READING WAS 55.7 DEGREES CENTIGRADE. THESE ARE FAILING RESULTS. THE THERMOMETER WILL BE SENT TO SUPPLIER FOR ROOT CAUSE ANALYSIS.

Description of Event or Problem · 1

A CONSUMER REPORTED THAT HER THERMOMETER HAD ALLEGEDLY GIVEN AN INACCURATE READING. THE DEVICE WAS READING IN THE NORMAL RANGE DESPITE THEIR SON HAVING A FEVER. THE CHILD WAS GIVEN MOTRIN (ORALLY) AND WAS SEEN BY THEIR PHYSICIAN ON THE FOLLOWING DAY. THERE WERE NO FURTHER COMPLICATIONS AND THE PT IS DOING WELL NOW.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
267853 BRAUN THERMOMETER FLL KAZ USA, INC. IRT-302 950

Patients

Seq Age Sex Outcome Treatment
1 UNK Other