FOOT CONTROL, SC2100
Report
- Report Number
- 1045834-2014-12036
- Event Type
- Malfunction
- Date Received
- May 9, 2014
- Date of Event
- April 24, 2014
- Report Date
- April 24, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- PK080802
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. A FUNCTIONAL ASSESSMENT REVEALED THAT THE CORD WAS DAMAGED. THEREFORE, THE REPORTED CONDITION WAS DUPLICATED AND CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE MISUSE/ABUSE (OR POSSIBLY USER ERROR) FROM PULLING THE DEVICE AROUND BY ITS CORD. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT DURING A EAR NOSE THROAT (ENT) SURGICAL PROCEDURE THE FOOT CONTROL DEVICE WOULD NOT FUNCTION. THERE WAS A REPORTED DELAY OF FIVE MINUTES TO THE PLANNED SURGICAL PROCEDURE. A SPARE IDENTICAL DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 280687 | FOOT CONTROL, SC2100 | MOTOR, DRILL, ELECTRIC - FOOT CONTROL | HBC | DEPUY SYNTHES POWER TOOLS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |