FDA Adverse Event Malfunction Summary report: N

FOOT CONTROL, SC2100

MDR report key: 3802941 · Received May 9, 2014

Report

Report Number
1045834-2014-12036
Event Type
Malfunction
Date Received
May 9, 2014
Date of Event
April 24, 2014
Report Date
April 24, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK080802
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. A FUNCTIONAL ASSESSMENT REVEALED THAT THE CORD WAS DAMAGED. THEREFORE, THE REPORTED CONDITION WAS DUPLICATED AND CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE MISUSE/ABUSE (OR POSSIBLY USER ERROR) FROM PULLING THE DEVICE AROUND BY ITS CORD. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A EAR NOSE THROAT (ENT) SURGICAL PROCEDURE THE FOOT CONTROL DEVICE WOULD NOT FUNCTION. THERE WAS A REPORTED DELAY OF FIVE MINUTES TO THE PLANNED SURGICAL PROCEDURE. A SPARE IDENTICAL DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
280687 FOOT CONTROL, SC2100 MOTOR, DRILL, ELECTRIC - FOOT CONTROL HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1