FDA Adverse Event Injury Summary report: N

EVITA, MODEL UNKNOWN

MDR report key: 3802866 · Received April 30, 2014

Report

Report Number
9611500-2014-00034
Event Type
Injury
Date Received
April 30, 2014
Date of Event
April 12, 2014
Report Date
April 30, 2014
Manufacturer
DRAEGER MEDICAL GMBH
Product Code
CBK
PMA / PMN Number
K961687
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DRAGER TRIED TO GET MORE DETAILED INFORMATION, BUT WITHOUT SUCCESS. IT COULD NOT BE CLARIFIED WHICH DEVICE WAS IN USE DURING THE EVENT. PROBABLY IT WAS A DEVICE OF EVITA FAMILY. AS THE CONCERNED DEVICE WAS NOT IDENTIFIED, NO DEVICE LOG WAS AVAILABLE FOR INVESTIGATION. THE CONCERNED CO2-CUVETTE WAS ALSO NOT MADE AVAILABLE. DUE TO THIS MISSING INFORMATION ONLY A GENERAL STATEMENT CAN BE GIVEN. TH ECOR-CUVETTE IS PLACED BETWEEN Y-PIECE AND ENDOTRACHEAL TUBE. AT THIS POSITION IT IS POSSIBLE TO DISTINGUISH THE VALUES FOR INSPIRATION AND EXPIRATION. IF THE CUVETTE DISCONNECTS AS REPORTED, THE COMPLETE HOSE SYSTEM IS DISCONNECTED FORM THE PATIENT. THIS WILL RESULT IN ALARMS OF HIGH PRIORITY WITH EACH INTENSIVE CARE VENTILATOR, AS VOLUME AND PRESSURE MONITORING ARE ALWAYS INTEGRATED FUNCTIONS OF THOSE DEVICES. MULTIPLE ALARMS FOR TIDAL VOLUME, PRESSURE, LEAKAGE AND CO2-APNOE (IF AVAILABLE) WILL FOLLOW, SEQUENCE DEPENDING ON MEASUREMENT RESULTS AND ALARM SETTINGS OF USERS. DEVICE RELATED INVESTIGATION WAS NOT POSSIBLE. NO DEVICE MALFUNCTION WAS REPORTED. THE REPORTED EVENT IS AN ISOLATED CASE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A CO2-CUVETTE DISCONNECTED DURING VENTILATION. CLINICAL CONSEQUENCE REPORTED: DESATURATION, BRADYCARDIA 57, EXTERNAL CARDIAC MASSAGE FOR 4 MINUTES + ADRENALINE, HEMODYNAMIC UNSTABLE, UNFAVORABLE EVOLUTION, DEATH OF THE PATIENT 6 HOURS AFTER INCIDENT. DEATH WAS EXPECTED DUE TO PATIENTS BAD STATE OF HEALTH AND NOT ATTRIBUTED TO THE INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
258234 EVITA, MODEL UNKNOWN VENTILATOR, CRITICAL CARE CBK DRAEGER MEDICAL GMBH UNK NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention