FDA Adverse Event Malfunction Summary report: N

ROTATING MULTIPLE CLIP APPLIER

MDR report key: 3802810 · Received May 9, 2014

Report

Report Number
3005075853-2014-03157
Event Type
Malfunction
Date Received
May 9, 2014
Date of Event
April 15, 2014
Report Date
April 15, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS OF THE ER420 DEVICE FOUND THAT IT WAS RECEIVED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE INCIDENT REPORTED THE DEVICE WAS FUNCTIONALLY EVALUATED. UPON FIRING OF THE DEVICE, THE REMAINING 17 CLIPS WERE NOT PROPERLY FED INTO THE JAWS, CAUSING THE CLIP TO BE FED SIDEWAYS AND EJECTED; FINALLY, IT LOCKED OUT AS INTENDED. IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE FOUND FEEDING ISSUE. HOWEVER AN INVESTIGATION HAS BEEN INITIATED TO ADDRESS THE POTENTIAL ROOT CAUSE OF THE DROPPING OR EJECTED CLIPS. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A BARIATRIC PROCEDURE, INCOMPLETE CLIP CLOSURE; CLIPS WERE MALFORMED. IT IS UNKNOWN HOW THE CASE WAS COMPLETED. THERE WERE NO PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
280543 ROTATING MULTIPLE CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. UNK J4CU9Y

Patients

Seq Age Sex Outcome Treatment
1