FDA Adverse Event Malfunction Summary report: N

SYMBIQ DUAL CHANNEL

MDR report key: 3802798 · Received February 21, 2014

Report

Report Number
9615050-2014-01384
Event Type
Malfunction
Date Received
February 21, 2014
Date of Event
January 1, 2014
Report Date
January 22, 2014
Manufacturer
HOSPIRA COSTA RICA LTD
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z-0070-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING THE DEVICE TOUCHSCREEN WAS FOUND TO BE OUT OF CALIBRATION. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT THE DEVICE TOUCHSCREEN WAS OUT OF CALIBRATION. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTED ADVERSE PATIENT EVENTS, NO NEED FOR MEDICAL INTERVENTION AND NO DELAYS IN CRITICAL THERAPY WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE TOUCHSCREEN WOULD NOT CALIBRATE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
108932 SYMBIQ DUAL CHANNEL 80FRN FRN HOSPIRA COSTA RICA LTD NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK