FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ DUAL CHAN INF
MDR report key: 3802793
·
Received February 21, 2014
Report
- Report Number
- 9615050-2014-01393
- Event Type
- Malfunction
- Date Received
- February 21, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 22, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- Z-0070-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING THE DEVICE DID NOT PASS TOUCHSCREEN TEST. THE PROBABLE CAUSE WAS FLUID INGRESS FROM THE CLEANING PRODUCTS USED AT THE USER FACILITY. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THE TOUCHSCREEN NEEDS CALIBRATING. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT PASS THE TOUCHSCREEN TEST. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 109022 | SYMBIQ DUAL CHAN INF | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |