PLUM A+ DRIVER ED 2
Report
- Report Number
- 9615050-2014-01404
- Event Type
- Malfunction
- Date Received
- February 21, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 28, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K042081
- Removal / Correction Number
- Z-0624-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
TESTING AND INVESTIGATION FOUND THAT THE DEVICE DOOR ROLLER WAS MISSING AND THE DOOR ROLLER PIN WAS BROKEN. FURTHER TESTING, FOUND THE DEVICE HAD UNRESTRICTED FLOW WHEN THE DOOR WAS OPENED. THIS WAS DUE TO THE MISSING DEVICE DOOR ROLLER AND BROKEN DOOR ROLLER PIN. THE MISSING DEVICE DOOR ROLLER AND BROKEN DOOR ROLLER PIN PREVENTED PROPER ACTUATION OF THE REGULATORY CLOSER ON THE DEVICE MECHANISM CAUSING UNRESTRICTED FLOW WHEN THE DOOR WAS OPENED. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THAT DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS BROKEN. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 109004 | PLUM A+ DRIVER ED 2 | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | PLUM A+ SOFTWARE MODULE: LIST #UNK, SN UNK |