FDA Adverse Event Malfunction Summary report: N

PLUM A+ DRIVER ED 2

MDR report key: 3802772 · Received February 21, 2014

Report

Report Number
9615050-2014-01404
Event Type
Malfunction
Date Received
February 21, 2014
Date of Event
January 1, 2014
Report Date
January 28, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K042081
Removal / Correction Number
Z-0624-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THAT THE DEVICE DOOR ROLLER WAS MISSING AND THE DOOR ROLLER PIN WAS BROKEN. FURTHER TESTING, FOUND THE DEVICE HAD UNRESTRICTED FLOW WHEN THE DOOR WAS OPENED. THIS WAS DUE TO THE MISSING DEVICE DOOR ROLLER AND BROKEN DOOR ROLLER PIN. THE MISSING DEVICE DOOR ROLLER AND BROKEN DOOR ROLLER PIN PREVENTED PROPER ACTUATION OF THE REGULATORY CLOSER ON THE DEVICE MECHANISM CAUSING UNRESTRICTED FLOW WHEN THE DOOR WAS OPENED. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS BROKEN. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
109004 PLUM A+ DRIVER ED 2 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA PLUM A+ SOFTWARE MODULE: LIST #UNK, SN UNK