HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-14979
- Event Type
- Malfunction
- Date Received
- May 9, 2014
- Date of Event
- April 14, 2014
- Report Date
- April 14, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN, A DEVICE ANALYSIS COULD NOT BE COMPLETED. HOWEVER, INITIATING THERAPY PRIOR TO CONNECTING TO THE DEVICE IS A KNOWN CAUSE OF THIS ALARM. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR WHEN AND HOW TO CONNECT TO THE DISPOSABLE SET AND INSTRUCTS THE USER TO CONNECT THE TRANSFER SET TO THE PATIENT LINE PRIOR TO STARTING THE INITIAL DRAIN. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SYSTEM ERROR (SE) 2240 ALARM (AIR IN LINE) OCCURRED ON A HOMECHOICE (HC) DEVICE DURING THE SECOND DWELL CYCLE OF AUTOMATED PERITONEAL DIALYSIS (APD) THERAPY. TROUBLESHOOTING REVEALED THAT THE HOME PATIENT (HP) PRESSED GO BEFORE CONNECTING TO THE DEVICE. THE HP THEN PROCEEDED TO CONNECT TO THE MACHINE. A TECHNICAL SERVICE REPRESENTATIVE (TSR) REVIEWED PROPER PROCEDURES WITH THE HP. THE TSR INSTRUCTED THE HP TO COMPLETE THERAPY WITH NEW SUPPLIES OR CONTINUOUS AMBULATORY PERITONEAL DIALYSIS (CAPD). THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 280800 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | HOMECHOICE |