FDA Adverse Event Malfunction Summary report: N

PLUM A+ PUMP REFURB

MDR report key: 3802760 · Received February 21, 2014

Report

Report Number
9615050-2014-01410
Event Type
Malfunction
Date Received
February 21, 2014
Date of Event
January 1, 2014
Report Date
January 22, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K070398
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RP
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION IT HAS NOT YET BEEN RECEIVED. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED DURING TESTING AT THE USER FACILITY THE DEVICE CONTINUED TO DELIVER AT THE PROGRAMMED RATE, AFTER THE PROGRAMMED VOLUME TO BE INFUSED WAS COMPLETE. THERE WAS NO REPORTS OF ANY ADVERSE PATIENT EVENTS, AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
109000 PLUM A+ PUMP REFURB 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA PLUM A+ SOFTWARE MODULE: LIST #12097,| SN (B)(4)