FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3802659
·
Received February 25, 2014
Report
- Report Number
- 1720753-2014-01783
- Event Type
- Malfunction
- Date Received
- February 25, 2014
- Date of Event
- February 7, 2014
- Report Date
- February 25, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE GENERATOR INTERFACE BOARD AND THE LEMO CONNECTOR WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THERE WAS A STATOR NOT ON ERROR MESSAGE. THIS ERROR WILL LIKELY CAUSE THE SYSTEM TO SHUT DOWN UN-COMMANDED. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 116747 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |