GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2014-00356
- Event Type
- Malfunction
- Date Received
- May 9, 2014
- Date of Event
- April 1, 2014
- Report Date
- April 2, 2014
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
- Product Code
- GEX
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER (B)(4): THE FIBER SHOWS A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE METAL CAP EXHIBITS MILD DETRITUS ADHESION; THE GLASS CAP EXHIBITS MODERATE DEVITRIFICATION AT THE OUTPUT WINDOW. BASED ON THE ANALYSIS ABOVE, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH LIMITING THE PERFORMANCE OF THE FIBER.
IT WAS REPORTED THAT DURING A PROSTATE PROCEDURE, @ 30396 JOULES OF USE, THE FIBER EXHIBITED THE AIMING BEAM AND VAPORIZATION STRAIGHT OUT OF THE FIBER. THE SECOND FIBER, @ 33582 JOULES OF USE, EXHIBITED SIGNIFICANTLY DIMINISHED VAPORIZATION. THE THIRD FIBER, @ 102375 JOULES OF USE, EXHIBITED THE AIMING BEAM FIRES STRAIGHT OUT OF THE FIBER. THE FOURTH FIBER, @ 34168 JOULES OF USE, EXHIBITED SIGNIFICANTLY DIMINISHED VAPORIZATION EFFICIENCY. THE PROCEDURE WAS COMPLETED WITH THE USE OF A FIFTH FIBER @ 8893 JOULES OF USE. ¿NO PATIENT INJURY¿ REPORTED. THIS REPORT IS FOR THE FOURTH FIBER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 281873 | GREENLIGHT MOXY FIBER OPTIC | POWERED LASER SURGICAL INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) | 0010-2400 | 349A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |