TPS HANDPIECE CORD
Report
- Report Number
- 0001811755-2014-01686
- Event Type
- Malfunction
- Date Received
- May 9, 2014
- Date of Event
- April 14, 2014
- Report Date
- April 14, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K943540
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
THE REPORTED EVENT WAS NOT CONFIRMED DURING THE DEVICE EVALUATION AND NO OTHER FAILURES WERE DETECTED. THE DEVICE WAS DISCARDED BY THE MANUFACTURER. IT WAS CONFIRMED THAT THE HANDPIECE (CORE IMPACTION DRILL 5400300000 ((B)(4))), WHICH IS LISTED HAD FAILURES THAT MAY HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. HOWEVER, AS THERE ARE NO HALL SENSORS IN THIS HANDPIECE, A BIAS-CURRENT WARNING SERVES NO PREVENTATIVE PURPOSE IN TERMS OF RUN-ON; THEREFORE, A REPORT WILL NOT BE FILED ON THE HANDPIECE.
A FOLLOW UP REPORT WILL BE FILED AFTER THE QUALITY INVESTIGATION HAS BEEN COMPLETED.
IT WAS REPORTED THAT DURING SET UP AT THE USER FACILITY THE TPS HANDPIECE CORD CAUSED A BIAS CURRENT MESSAGE TO BE DISPLAYED ON THE CONSOLE, SIGNALING A CONDITION OCCURRED IN WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
IT WAS REPORTED THAT DURING SET UP AT THE USER FACILITY THE TPS HANDPIECE CORD CAUSED A BIAS CURRENT MESSAGE TO BE DISPLAYED ON THE CONSOLE, SIGNALING A CONDITION OCCURRED IN WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 281914 | TPS HANDPIECE CORD | DRIVER, WIRE, AND BONE DRILL, MANUAL | DZJ | STRYKER INSTRUMENTS-KALAMAZOO | 13070 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CORE CONSOLE 5400050000 ((B)(4))| CORE IMPACTION DRILL 5400300000 ((B)(4)) |