FDA Adverse Event Malfunction Summary report: N

ACCENT DR RF

MDR report key: 3802539 · Received May 9, 2014

Report

Report Number
2017865-2014-13758
Event Type
Malfunction
Date Received
May 9, 2014
Date of Event
March 17, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
DXY
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

NEW INFORMATION RECEIVED ON (B)(6) 2014 NOTES THAT THE PATIENT WAS HAVING AN ABLATION PROCEDURE, THE PULSE GENERATOR TRIPPED ELECTIVE REPLACEMENT INDICATOR DUE TO BEING EXPOSED TO ELECTROCAUTERY. AFTER A PRODUCT CODE DOWNLOAD, THE DEVICE RESUMED NORMALLY FUNCTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PULSE GENERATOR EXHIBITED PREMATURE ELECTIVE REPLACEMENT INDICATOR. AFTER A PRODUCT CODE DOWNLOAD THE DEVICE RESUMED NORMAL FUNCTION. THE PATIENT WOULD BE MONITORED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281836 ACCENT DR RF IMPLANTABLE PACEMAKER PULSE GENERATOR DXY ST. JUDE MEDICAL, INC., CRMD PM2210 NA

Patients

Seq Age Sex Outcome Treatment
1 76 YR