FDA Adverse Event
Malfunction
Summary report: N
ACCENT DR RF
MDR report key: 3802539
·
Received May 9, 2014
Report
- Report Number
- 2017865-2014-13758
- Event Type
- Malfunction
- Date Received
- May 9, 2014
- Date of Event
- March 17, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- DXY
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
NEW INFORMATION RECEIVED ON (B)(6) 2014 NOTES THAT THE PATIENT WAS HAVING AN ABLATION PROCEDURE, THE PULSE GENERATOR TRIPPED ELECTIVE REPLACEMENT INDICATOR DUE TO BEING EXPOSED TO ELECTROCAUTERY. AFTER A PRODUCT CODE DOWNLOAD, THE DEVICE RESUMED NORMALLY FUNCTION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PULSE GENERATOR EXHIBITED PREMATURE ELECTIVE REPLACEMENT INDICATOR. AFTER A PRODUCT CODE DOWNLOAD THE DEVICE RESUMED NORMAL FUNCTION. THE PATIENT WOULD BE MONITORED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 281836 | ACCENT DR RF | IMPLANTABLE PACEMAKER PULSE GENERATOR | DXY | ST. JUDE MEDICAL, INC., CRMD | PM2210 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR |