ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2014-03149
- Event Type
- Malfunction
- Date Received
- May 9, 2014
- Report Date
- March 13, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE DEVICE WAS RETURNED IN GOOD PHYSICAL CONDITION. THE DEVICE WAS FUNCTIONALLY TESTED ON THE GENERATOR AND THE HAND CONTROL BUTTONS WERE NOT FUNCTIONAL. HOWEVER, THE DEVICE DID ACTIVATE WHEN TESTED WITH THE FOOT SWITCH. NO ALERT SCREENS WERE DISPLAYED ON GENERATOR. THE DEVICE WAS DISASSEMBLED TO INSPECT INTERNAL COMPONENTS. CORROSION WAS FOUND AT THE HAND ACTIVATION DOMES. THE CORROSION IN THE DOMES IS POSSIBLY CAUSED WHEN THE DEVICE WAS SOAKED FOR RE-USE OR THE DEVICE BEING SOAKED FOR CLEANING BEFORE SHIPMENT FOR ANALYSIS. THE ROOT CAUSE CANNOT BE IDENTIFIED BECAUSE IT IS UNKNOWN WHEN THE CORROSION OCCURRED. THE CORROSION WOULD HAVE INHIBITED ELECTRICAL CONTACT BETWEEN THE DOME AND THE CIRCUIT. OUR MANUFACTURING, STERILIZATION, PACKAGING AND SHIPMENT PROCESSES DO NOT INTRODUCE CORROSION TO THE DEVICE. IT IS PROBABLE THAT THIS MAY HAVE AFFECTED THE FUNCTIONALITY OF THE HAND ACTIVATION BUTTONS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC SIGMA PROCEDURE GENERATOR SHOWED ERROR CODE: ¿REDUCE PRESSURE ON BLADE¿; ¿REPLACE INSTRUMENT¿; WITH GEN11. NO FURTHER INFORMATION IS AVAILABLE. NO PATIENT CONSEQUENCES REPORTED. PROCEDURE WAS COMPLETED WITH SAME LIKE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 281315 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR, HANDPIECE |