FDA Adverse Event Malfunction Summary report: N

INCANDESCENT LAMP

MDR report key: 38023 · Received September 16, 1996

Report

Report Number
MW4001490
Event Type
Malfunction
Date Received
September 16, 1996
Report Date
May 17, 1996
Manufacturer
BULBTRONICS
Product Code
EAZ
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

SUSPECTED MFG DEFECT. THESE LAMPS ARE IDENTIFIED AS REPLACEMENT BULBS FOR A SPECIFIC BRAND DENTAL OPERATING LIGHTS. HOWEVER, THE LAMPS ARE OF INSUFFICIENT LENGTH TO MAKE FIRM CONTACT WITH THE LAMP HOLDER SOCKETS WHICH CREATES AN INTERMITTENT CONTINUITY CONDITION. THIS LEADS TO ELECTRICAL ARCING AND SUBSEQUENT CARBON BUILD UP ON THE SOCKET AND BULB CONTACT SURFACES. THIS CONDITION REQUIRES REPLACEMENT OF BOTH THE SOCKETS AND BULB. ADDING TO THE PROBLEM IS POOR QUALITY FINISH ON THE BULB ENDS WHICH HASTEN THE CARBON BUILD UP. WORK ORDER REVIEW INDICATES 18 SOCKETS WERE REPLACED DUE TO THIS CONDITION, AT $ 27.00 EACH, IN THE LAST SIX MONTHS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INCANDESCENT LAMP INCANDESCENT LAMP EAZ BULBTRONICS * *

Patients

Seq Age Sex Outcome Treatment
1 *