FDA Adverse Event Malfunction Summary report: N

2008K HEMODIALYSIS SYS. OLC/DIASAFE PLUS

MDR report key: 3802283 · Received March 24, 2014

Report

Report Number
2937457-2014-00433
Event Type
Malfunction
Date Received
March 24, 2014
Date of Event
February 22, 2014
Report Date
February 26, 2014
Manufacturer
FRESENIUS MEDICAL CARE NORTH AMERICA
Product Code
KDI
PMA / PMN Number
K994267
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATIONS FINDINGS TO DATE INDICATED THE REPORTED MALFUNCTION OCCURRED DURING RECIRCULATION AND PRIME (MACHINE SET-UP), AND NOT DURING DIALYSIS MODE. THE USER VISUALLY OBSERVED THE SALINE BAG REFILLING WITH DIALYSATE DURING CIRCULATION. THERE HAVE BEEN NO ADVERSE EVENTS ASSOCIATED WITH THE REPORTED ISSUE. THE REPORT IS BEING INVESTIGATED BY THE MFR VIA A CAPA. THE INVESTIGATION IS PENDING A SUPPLEMENTAL MDR WILL BE FILED AT THE COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

THE USER FACILITY REPORTED A SALINE BAG BACK FILLED DURING RECIRCULATION MODE. A PT WAS NOT CONNECTED TO THE MACHINE AT THE TIME OF THE INCIDENT. ADDITIONAL ATTEMPTS TO THE CUSTOMER HAVE BEEN MADE WITH NO ADDITIONAL INFO PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
171340 2008K HEMODIALYSIS SYS. OLC/DIASAFE PLUS KDI FRESENIUS MEDICAL CARE NORTH AMERICA

Patients

Seq Age Sex Outcome Treatment
1 UNK SALINE