2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS
Report
- Report Number
- 2937457-2014-00436
- Event Type
- Malfunction
- Date Received
- March 24, 2014
- Date of Event
- March 10, 2014
- Report Date
- March 10, 2014
- Manufacturer
- FRESENIUS MEDICAL CARE NORTH AMERICA
- Product Code
- KDI
- PMA / PMN Number
- K994267
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATIONS FINDINGS TO DATE INDICATED THE REPORTED MALFUNCTION OCCURRED DURING RECIRCULATION AND PRIM (MACHINE SET-UP), AND NOT DURING DIALYSIS MODE. THE USER VISUALLY OBSERVED THE SALINE BAG REFILLING WITH DIALYSATE DURING CIRCULATION. THERE HAVE BEEN NO ADVERSE EVENTS ASSOCIATED WITH THE REPORTED ISSUE. THE REPORT IS BEING INVESTIGATED BY THE MANUFACTURER VIA A CAPA. THE INVESTIGATION IS PENDING A SUPPLEMENTAL MDR WILL BE FILED AT THE COMPLETION OF THE INVESTIGATION.
A USER FACILITY REPORTED A SALINE BAG BACK FILLED DURING RECIRCULATION MODE. A PATIENT WAS NOT CONNECTED TO THE MACHINE AT THE TIME OF THE INCIDENT. THERE WERE 2 LEAKING VALUES. THE FACILITY TECHNICIAN ORDERED A FLOW REGULATOR AND LOADING PRESSURE REGULATOR TO REPLACE BUT PARTS WERE ON BACK ORDER. MACHINE HAS BEEN OUT OF SERVICE. TECHNICIAN WILL REPLACE THE FLOW REGULATOR AND LOADING PRESSURE REGULATORS WHEN PARTS ARE RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 171274 | 2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS | KDI | FRESENIUS MEDICAL CARE NORTH AMERICA | K2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK SALINE |