FDA Adverse Event Malfunction Summary report: N

2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS

MDR report key: 3802281 · Received March 24, 2014

Report

Report Number
2937457-2014-00436
Event Type
Malfunction
Date Received
March 24, 2014
Date of Event
March 10, 2014
Report Date
March 10, 2014
Manufacturer
FRESENIUS MEDICAL CARE NORTH AMERICA
Product Code
KDI
PMA / PMN Number
K994267
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATIONS FINDINGS TO DATE INDICATED THE REPORTED MALFUNCTION OCCURRED DURING RECIRCULATION AND PRIM (MACHINE SET-UP), AND NOT DURING DIALYSIS MODE. THE USER VISUALLY OBSERVED THE SALINE BAG REFILLING WITH DIALYSATE DURING CIRCULATION. THERE HAVE BEEN NO ADVERSE EVENTS ASSOCIATED WITH THE REPORTED ISSUE. THE REPORT IS BEING INVESTIGATED BY THE MANUFACTURER VIA A CAPA. THE INVESTIGATION IS PENDING A SUPPLEMENTAL MDR WILL BE FILED AT THE COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

A USER FACILITY REPORTED A SALINE BAG BACK FILLED DURING RECIRCULATION MODE. A PATIENT WAS NOT CONNECTED TO THE MACHINE AT THE TIME OF THE INCIDENT. THERE WERE 2 LEAKING VALUES. THE FACILITY TECHNICIAN ORDERED A FLOW REGULATOR AND LOADING PRESSURE REGULATOR TO REPLACE BUT PARTS WERE ON BACK ORDER. MACHINE HAS BEEN OUT OF SERVICE. TECHNICIAN WILL REPLACE THE FLOW REGULATOR AND LOADING PRESSURE REGULATORS WHEN PARTS ARE RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
171274 2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS KDI FRESENIUS MEDICAL CARE NORTH AMERICA K2

Patients

Seq Age Sex Outcome Treatment
1 UNK SALINE