ECHELON LINEAR CUTTER
Report
- Report Number
- 3005075853-2014-03146
- Event Type
- Malfunction
- Date Received
- May 9, 2014
- Report Date
- April 14, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K081146
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). THE ANALYSIS FOUND THAT ONE EC60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH ONE ECR60W CARTRIDGE LOADED IN THE DEVICE. THE CARTRIDGE RELOAD WAS RECEIVED FULLY FIRED AND IN GOOD VISUAL CONDITION. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. IT SHOULD BE NOTED THAT IN ORDER TO OPEN A DEVICE THAT HAS BEEN PARTIALLY FIRED OR FULLY FIRED A REVERSE STROKE NEEDS TO BE PERFORMED TRIGGER TO TRIGGER TO HANDLE IN ORDER TO RETURN THE KNIFE TO THE HOME POSITION (INDICATOR IN THE "0" POSITION) AND PRESS THE ANVIL RELEASE BUTTON TO OPEN. EVENT COULD NOT BE CONFIRMED AS THE DEVICE CLOSED AND OPENED WITHOUT ANY DIFFICULTIES NOTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE COULD NOT BE OPENED AFTER INSERTING THROUGH A TROCAR. NO FURTHER INFORMATION IS AVAILABLE. NO PATIENT CONSEQUENCES REPORTED. PROCEDURE WAS COMPLETED WITH SAME LIKE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 280312 | ECHELON LINEAR CUTTER | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | K4DT91 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |