GREENLIGHT MOXY FIBER OPTIC

MDR report key: 3802103 · Received May 9, 2014

Report

Report Number
2937094-2014-00365
Event Type
Malfunction
Date Received
May 9, 2014
Date of Event
April 3, 2014
Report Date
April 4, 2014
Manufacturer
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
Product Code
GEX
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER MODEL # 0010-2400; LOT #345A; SERIAL # (B)(4): THE FIBER SHOWS A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE METAL CAP EXHIBITS MODERATE CHAR AND DETRITUS ADHESION; THE GLASS CAP EXHIBITS MODERATE DEVITRIFICATION AT THE OUTPUT WINDOW. BASED ON THE ANALYSIS ABOVE, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH LIMITING THE PERFORMANCE OF THE FIBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE USING THE SIDE-FIRING SURGICAL FIBER DURING A PROSTATE PROCEDURE, THE FIBER HAD A "MISFUNCTION ERROR UNUSABLE." THE PROCEDURE WAS COMPLETED USING A SECOND FIBER. PATIENT OUTCOME: "NO DAMAGES TO THE PATIENT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
280514 GREENLIGHT MOXY FIBER OPTIC POWERED LASER SURGICAL INSTRUMENT GEX AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) 0010-2400 345A

Patients

Seq Age Sex Outcome Treatment
1