FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE (ITEM # UNK)

MDR report key: 3801953 · Received February 14, 2014

Report

Report Number
1419937-2014-00177
Event Type
Malfunction
Date Received
February 14, 2014
Date of Event
January 24, 2014
Report Date
February 5, 2014
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

IN F/U WITH THE CUSTOMER SHE STATED THAT SHE HAS THROWN THE OLD ADAPTER AWAY AND PURCHASE A NEW ONE. THE CUSTOMER ALSO STATED THAT THERE WERE NO INJURIES DUE TO THIS ISSUE. THIS ISSUE WITH THE DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FAIL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORESEEABLE MISUSE. THE PACKAGING USED BY THE MANUFACTURER TO SHIP THE TRANSFORMERS TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING, HANDLING, AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STACKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PROTECT THE TRANSFORMERS DURING SHIPPING.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO CUSTOMER SERVICE THAT HER WALL PLUG CRACKED IN HALF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
97798 PUMP IN STYLE (ITEM # UNK) HGX MEDELA, INC. 57XXXX/9207010 UNK

Patients

Seq Age Sex Outcome Treatment
1