FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3801952 · Received February 4, 2014

Report

Report Number
1720753-2014-01150
Event Type
Malfunction
Date Received
February 4, 2014
Date of Event
January 15, 2014
Report Date
February 4, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE MAINFRAME SYSTEM BATTERIES WAS EVALUATED AND REPLACED. THE HV CABLE CONNECTIONS WERE ALSO VAPOR PROOFED AND RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WAS AN 'SATURATION FAULT ERROR' MESSAGE DISPLAYED. THIS ERROR IS LIKELY TO RESULT IN AN IMMEDIATE LOSS OF SYSTEM FUNCTIONALITY AND AN UN-COMMANDED SHUT DOWN. THERE WAS NO REPORT OF A PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.