VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910) SUTURE
Report
- Report Number
- 2210968-2014-05924
- Event Type
- Malfunction
- Date Received
- May 9, 2014
- Date of Event
- March 7, 2014
- Report Date
- March 17, 2014
- Manufacturer
- ETHICON INC.
- Product Code
- GAM
- PMA / PMN Number
- K032420
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
CONCLUSION: THE ACTUAL NEEDLE REVEALED INDENTS AND SCUFF MARKS AROUND THE BREAK AREA PRODUCED DURING HANDLING BY THE SURGICAL NEEDLE HOLDERS OR SOME OTHER GRIPPING DEVICE. THE NEEDLE FRACTURED DUE TO TENSILE OVERLOAD GENERATED DURING SEVERE MECHANICAL DEFORMATION, WITH SIGNS OF DUCTILE FAILURE. THERE ARE NO SIGNS OF MANUFACTURING DEFECTS FOUND ON THE NEEDLE. THE DEVICE INFORMATION FOR USE CAUTIONS THE USER THAT "TO AVOID DAMAGING NEEDLE POINTS AND SWAGE AREAS, GRASP THE NEEDLE IN AN AREA ONE-THIRD (1/3) TO ONE-HALF (1/2) OF THE DISTANCE FROM THE SWAGED END TO THE POINT."
(B)(4). CONCLUSION: TO DATE, THE DEVICE HAS NOT BEEN RETURNED. IF THE PRODUCT IS RETURNED FOR EVALUATION, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2014 AND SUTURE WAS USED. DURING THE FIRST TISSUE PASS, THE NEEDLE BENT AND THEN BROKE AT THE SWAGE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.