FDA Adverse Event Malfunction Summary report: N

VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910) SUTURE

MDR report key: 3801926 · Received May 9, 2014

Report

Report Number
2210968-2014-05924
Event Type
Malfunction
Date Received
May 9, 2014
Date of Event
March 7, 2014
Report Date
March 17, 2014
Manufacturer
ETHICON INC.
Product Code
GAM
PMA / PMN Number
K032420
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONCLUSION: THE ACTUAL NEEDLE REVEALED INDENTS AND SCUFF MARKS AROUND THE BREAK AREA PRODUCED DURING HANDLING BY THE SURGICAL NEEDLE HOLDERS OR SOME OTHER GRIPPING DEVICE. THE NEEDLE FRACTURED DUE TO TENSILE OVERLOAD GENERATED DURING SEVERE MECHANICAL DEFORMATION, WITH SIGNS OF DUCTILE FAILURE. THERE ARE NO SIGNS OF MANUFACTURING DEFECTS FOUND ON THE NEEDLE. THE DEVICE INFORMATION FOR USE CAUTIONS THE USER THAT "TO AVOID DAMAGING NEEDLE POINTS AND SWAGE AREAS, GRASP THE NEEDLE IN AN AREA ONE-THIRD (1/3) TO ONE-HALF (1/2) OF THE DISTANCE FROM THE SWAGED END TO THE POINT."

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: TO DATE, THE DEVICE HAS NOT BEEN RETURNED. IF THE PRODUCT IS RETURNED FOR EVALUATION, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2014 AND SUTURE WAS USED. DURING THE FIRST TISSUE PASS, THE NEEDLE BENT AND THEN BROKE AT THE SWAGE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281618 VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910) SUTURE SUTURE, ABSORBABLE, SYNTHETIC GAM ETHICON INC. GL2580

Patients

Seq Age Sex Outcome Treatment
1