FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 3801837 · Received February 11, 2014

Report

Report Number
1720753-2014-01389
Event Type
Malfunction
Date Received
February 11, 2014
Date of Event
January 22, 2014
Report Date
February 11, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE AUXILIARY INTERFACE PCB WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE FE REPORTED THAT THE SYSTEM INTERMITTENTLY HAD FAILURES BOOTING UP AND PERFORMING FLUOROSCOPY. THIS RESULTS INTERMITTENT, RECOVERABLE LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
90118 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1