FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3801781
·
Received January 8, 2014
Report
- Report Number
- 1314492-2014-01304
- Event Type
- Malfunction
- Date Received
- January 8, 2014
- Date of Event
- November 1, 2013
- Report Date
- December 11, 2013
- Manufacturer
- BAXTER HLTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF COMPLAINT (B)(4). THE "AIR-IN-LINE" ALARMS WERE CONFIRMED THROUGH THE EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADD'L INFO IS RECEIVED, A FOLLOW-UP WILL BE SENT. THE ULTRASONIC PUSHER WAS REPLACED.
Description of Event or Problem · 1
DURING THE EVALUATION OF A RETURNED SPECTRUM INFUSION PUMP, 4 OCCURRENCES OF "AIR-IN-LINE" ALARMS WERE IDENTIFIED IN THE EVENT HISTORY LOG. ANY PATIENT INJURY OR MEDICAL INTERVENTION IS UNK SINCE THESE ITEMS WERE FOUND DURING EVALUATION OF THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 13548 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HLTHCARE CORP | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |