SPECTRUM INFUSION PUMP
Report
- Report Number
- 1314492-2014-01320
- Event Type
- Malfunction
- Date Received
- January 8, 2014
- Date of Event
- November 1, 2013
- Report Date
- December 11, 2013
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF COMPLAINT (B)(4). THE REPORTED "DOWNSTREAM OCCLUSION" ALARMS WERE CONFIRMED THROUGH THE EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT. THE FORCE SENSOR ASSEMBLY, FORCE SENSOR GASKET, FORCE SENSOR PUSHER, DOWNSTREAM TUBING GUIDE, AND INPUT/OUTPUT PRINTED CIRCUIT BOARD WERE REPLACED.
DURING THE EVAL OF A RETURNED SPECTRUM INFUSION PUMP, 217 OCCURRENCES OF "DOWNSTREAM OCCLUSION" ALARMS WERE IDENTIFIED IN THE EVENT HISTORY LOG. ANY PT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION IS UNK SINCE THESE ITEMS WERE FOUND DURING EVAL OF THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.