840 VENTILATOR
MDR report key: 3801708
·
Received February 11, 2014
Report
- Report Number
- 8020893-2014-00341
- Event Type
- Malfunction
- Date Received
- February 11, 2014
- Date of Event
- January 13, 2014
- Report Date
- January 13, 2014
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT IRELAND
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB) AND BACKLIGHT INVERTER PRINTED CIRCUIT BOARDS (PCB). THE UNIT PASSED EXTENDED SELF-TESTING.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD AN ERRATIC UPPER DISPLAY. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 88861 | 840 VENTILATOR | CBK: VENTILATOR, CONTINUOUS, FACILITY USE | CBK | COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT IRELAND | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |