FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 3801594 · Received May 9, 2014

Report

Report Number
2939301-2014-10769
Event Type
Malfunction
Date Received
May 9, 2014
Report Date
May 1, 2014
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ¿ (07/15/2014).THE PATIENT¿S TEST STRIPS HAVE BEEN RETURNED ON 6/9/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 7/3/2014 WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE TEST STRIPS WERE EVALUATED AND THE PRIMARY COMPLAINT WAS NOT CONFIRMED; HOWEVER, A SECONDARY ISSUE WAS NOTED WHERE ERROR 4 WAS OBSERVED DURING CONTROL SOLUTION TESTS. IN ADDITION, A TERTIARY ISSUE WAS NOTED WHEN THE TEST STRIPS WERE FOUND TO HAVE RESULTS ABOVE AND BELOW RANGE WHEN TESTED WITH CONTROL SOLUTION. QUATERNARY, THE RETURNED TEST STRIPS WERE FOUND TO HAVE SCRATCHED ELECTRODES WITH NO ASSIGNABLE CAUSE. QUINARY, THE METER WAS MISCLASSIFYING BLOOD AS ¿CONTROL SOLUTION¿ WHEN APPLIED TO A STRIP. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING ERROR 2. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
280824 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3584229

Patients

Seq Age Sex Outcome Treatment
1 69 YR