ANIMAS VIBE
Report
- Report Number
- 2531779-2014-13023
- Event Type
- Malfunction
- Date Received
- May 8, 2014
- Report Date
- April 24, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 07/17/2014.ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/30/2014 WITH THE FOLLOWING FINDINGS: REVIEW OF THE BLACK BOX AND ALARM HISTORY REVEALED CONSECUTIVE CALL SERVICE ALARMS RECORDED ON (B)(6) 2014. ON INVESTIGATION, THE PUMP WAS EXERCISED FOR 24 HOURS WITHOUT CALL SERVICE ALARMS OCCURRING. THE PUMP WAS FOUND TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION.
(B)(6). THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE DISTRIBUTOR CONTACTED ANIMAS ON (B)(6) 2014 ALLEGING THE PUMP EMITTED A CALL SERVICE ALARM THREE OR MORE TIMES WITHIN A 30-DAY PERIOD. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED ISSUE REMAINED UNRESOLVED WITH TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 279448 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |