ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-13016
- Event Type
- Malfunction
- Date Received
- May 8, 2014
- Report Date
- April 30, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PATIENT
Narratives
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. VISUAL INSPECTION OF THE PUMP FOUND NO COSMETIC DAMAGES. REVIEW OF BLACK BOX DATA DID NOT FIND ANY RELATED WARNINGS. THE BATTERY COMPARTMENT AND CAP WERE BOTH UNDAMAGED WITH NO EVIDENCE OF MOISTURE INTRUSION. THE PUMP POWERED UP AND FUNCTIONED PROPERLY. A REWIND/LOAD/PRIME SEQUENCE AND A 24-HOUR PUMP EXERCISE WERE EXECUTED WITHOUT INCIDENTS. THE ELECTRICAL CURRENT DRAWS WERE WITHIN SPECIFICATIONS. THE PUMP CASING WAS OPENED AND NO INTERNAL DEFECTS TO THE POWER CIRCUIT OR EVIDENCE OF MOISTURE INTRUSION WERE FOUND. THE INVESTIGATION WAS UNABLE TO VERIFY OR DUPLICATE THE REPORTED PUMP OVERHEATING ISSUE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS ALLEGING THAT THE PUMP AND THE BATTERY WERE HOT TO THE TOUCH. THE PATIENT STATED THAT THERE WAS NO LOSS OF POWER AND THERE WAS NO DAMAGE OR EVIDENCE OF MOISTURE INTRUSION IN THE PUMP. THE PATIENT DENIED ANY INJURY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 279421 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR |