FDA Adverse Event
Malfunction
Summary report: N
RIATA ACTIVE FIXATION
MDR report key: 3801391
·
Received May 8, 2014
Report
- Report Number
- 2938836-2014-10945
- Event Type
- Malfunction
- Date Received
- May 8, 2014
- Date of Event
- March 31, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
IT WAS REPORTED THAT AN ASYMPTOMATIC PATIENT WHO HAD BEEN LOST TO FOLLOW UP PRESENT IN HOSPITAL WITH HIGH PACING LEAD IMPEDANCE AND INCREASED CAPTURE THRESHOLD. THE DEVICE WAS REPROGRAMMED. AT A SUBSEQUENT FOLLOW UP, EXTERNALIZED CONDUCTORS WAS OBSERVED UNDER X-RAY. THE PATIENT IS DOING FINE AND WILL BE MONITORED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 278953 | RIATA ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 1580/65 | NA |