FDA Adverse Event
Injury
Summary report: N
UNIFY CRT-D, DF4 CONNECTOR
MDR report key: 3801327
·
Received May 8, 2014
Report
- Report Number
- 2938836-2014-10994
- Event Type
- Injury
- Date Received
- May 8, 2014
- Date of Event
- March 11, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NIK
- PMA / PMN Number
- P030054
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION: THE REPORTED FIELD EVENT OF NOISE ON WAS NOT CONFIRMED IN THE LABORATORY. THE DEVICE WAS TESTED ON THE BENCH AND USING AUTOMATED TESTING EQUIPMENT AND NO ANOMALIES WERE FOUND. THE CAUSE OF THE NOISE WAS NOT DETERMINED.
Description of Event or Problem · 1
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE DEVICE WAS ELECTIVELY REMOVED, AND THE INITIAL COMPLAINT OF NOISE WAS ONLY AGAINST THE ASSOCIATED RV LEAD.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WAS EXPLANTED AND REPLACED DUE TO NOISE. PATIENT CONDITION WAS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 279211 | UNIFY CRT-D, DF4 CONNECTOR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | NIK | ST. JUDE MEDICAL, INC., CRMD | CD3231-40Q | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR | Required Intervention | (B)(4) |