FDA Adverse Event Injury Summary report: N

UNIFY CRT-D, DF4 CONNECTOR

MDR report key: 3801327 · Received May 8, 2014

Report

Report Number
2938836-2014-10994
Event Type
Injury
Date Received
May 8, 2014
Date of Event
March 11, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
P030054
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION: THE REPORTED FIELD EVENT OF NOISE ON WAS NOT CONFIRMED IN THE LABORATORY. THE DEVICE WAS TESTED ON THE BENCH AND USING AUTOMATED TESTING EQUIPMENT AND NO ANOMALIES WERE FOUND. THE CAUSE OF THE NOISE WAS NOT DETERMINED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE DEVICE WAS ELECTIVELY REMOVED, AND THE INITIAL COMPLAINT OF NOISE WAS ONLY AGAINST THE ASSOCIATED RV LEAD.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS EXPLANTED AND REPLACED DUE TO NOISE. PATIENT CONDITION WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
279211 UNIFY CRT-D, DF4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC., CRMD CD3231-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 43 YR Required Intervention (B)(4)