PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-84469
- Event Type
- Malfunction
- Date Received
- May 8, 2014
- Date of Event
- May 5, 2014
- Report Date
- May 6, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. MEDTRONIC
IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP DID NOT GET THE NO DELIVERY ALARM AND THE BATTERY LASTED ONLY FOR A FEW DAYS. THE CUSTOMER EXPERIENCED HIGH BLOOD GLUCOSE OF 15.7MMOL/L AND HAD LOST CONFIDENCE IN THE DEVICE. THE CALLER MENTIONED THAT THE INFUSION SET WAS CHANGED THREE TIMES. ASSISTED THE CUSTOMER TO RUN THE HIGH PRESSURE TEST AND THE TEST FAILED TWICE. INITIALLY, IT WAS REPORTED THAT THE DEVICE WAS DOING WEIRD THINGS WHEN BOLUSING. THE CALLER STATE THAT THEY CALLED FROM THE SCHOOL AND THE CUSTOMER'S BLOOD GLUCOSE WAS 27MMOL/L. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 278645 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-754CAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |