FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3801217
·
Received May 8, 2014
Report
- Report Number
- 3004209178-2014-84476
- Event Type
- Malfunction
- Date Received
- May 8, 2014
- Date of Event
- March 14, 2014
- Report Date
- April 23, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP FAILED THE DISPLACEMENT TEST DURING REWIND, MOTOR ERROR AND OTHER ERROR ALARM FOUND IN THE ALARM HISTORY AS A RESULT OF A MOTOR ENCODER SIGNAL BEING OUT OF PHASE. THE DEVICE WAS RECEIVED WITH CRACKED RESERVOIR TUBE LIP AND SCRATCHED SCREEN.
Description of Event or Problem · 1
THE CUSTOMER REPORTED GETTING AN ERROR ALARM AND THE DEVICE ASKED TO REPROGRAM. THE CALLER STATED THAT THE INSULIN PUMP WAS STUCK AT TIME AND DATE SET UP. THE CUSTOMER STATED THAT THE DEVICE WAS EXPOSED TO A HIGH MAGNETIC FIELD. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 279026 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR |