FDA Adverse Event Malfunction Summary report: N

CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL PERICARDIAL BIOPROSTHESIS

MDR report key: 3801169 · Received May 8, 2014

Report

Report Number
2015691-2014-01078
Event Type
Malfunction
Date Received
May 8, 2014
Date of Event
April 4, 2014
Report Date
April 11, 2014
Manufacturer
EDWARDS LIFESCIENCES
Product Code
DYE
PMA / PMN Number
P860057/S029
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL ECHO REVIEW INFORMATION WAS RECEIVED.

Additional Manufacturer Narrative · 1

ADDITIONAL MANUFACTURER NARRATIVE: IT WAS REPORTED THAT THIS PATIENT EXPERIENCES REGURGITATION. THE TYPE AND CAUSE OF REGURGITATION VARIES DEPENDING UPON MULTIPLE FACTORS. OFTEN MODERATE TO HIGH REGURGITATION REQUIRING RE-OPERATION IN THE IMMEDIATE POST-OPERATIVE PERIOD IS DUE TO PROCEDURAL RELATED ISSUES AND IS UNRELATED TO THE DEVICE. HOWEVER, ADVANCES IN VALVE DESIGN AND BIOPROSTHETIC MATERIAL HAVE BEEN MADE WITH THE INTENTION OF REDUCING PERIVALVULAR OR CENTRAL LEAKS BY PROVIDING MORE EFFICIENT HEMODYNAMICS AND LONGER TISSUE DURABILITY. THE DEVICE REMAINS IMPLANTED IN THE PATIENT. THERE IS INSUFFICIENT INFORMATION TO CONCLUSIVELY DETERMINE THE ROOT CAUSE OF THIS EVENT. NO FURTHER ACTIONS ARE POSSIBLE WITH THE AVAILABLE INFORMATION. A SUPPLEMENTAL MDR WILL BE SUBMITTED IF ANY NEW INFORMATION IS OBTAINED.

Description of Event or Problem · 1

EDWARDS HAS RECEIVED INFORMATION OF A MITRAL BIOPROSTHETIC VALVE EXHIBITED MODERATE REGURGITATION ELEVEN (11) DAYS AFTER IMPLANT AND SEVERE REGURGITATION APPROXIMATELY NINE (9) MONTHS POST-IMPLANT. THE DEVICE REMAINS IMPLANTED. THERE ARE NO ADVERSE EVENTS REPORTED.

Description of Event or Problem · 1

ECHO REVIEW DONE BY INDEPENDENT CONSULTANT CONFIRMED RIGID LEAFLETS, MR 0 >>MR 2 AND REMNANT CORD IN LEFT VENTRICLE. LIKELY ABNORMAL SINCE IMPLANT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
279336 CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL PERICARDIAL BIOPROSTHESIS REPLACEMENT HEART VALVE DYE EDWARDS LIFESCIENCES 7000TFX

Patients

Seq Age Sex Outcome Treatment
1 69 YR Other