FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPSINSTRUMENT

MDR report key: 3801113 · Received May 8, 2014

Report

Report Number
2955842-2014-02882
Event Type
Malfunction
Date Received
May 8, 2014
Date of Event
April 9, 2014
Report Date
April 11, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND THE PITCH CABLE FRAYED AT THE DISTAL CLEVIS HUB. FRAYED STRANDS STUCK OUT AT THE INSTRUMENT'S WRIST. OTHER CABLES AT THE WRIST WERE NOT DAMAGED. ELECTRICAL CONTINUITY TESTING WAS PERFORMED AND PASSED. NO OTHER DAMAGE WAS FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT IN CENTRAL PROCESSING, A WIRE WAS SEPARATED AND STICKING OUT ON THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT. NOTHING REPORTEDLY FELL INTO A PATIENT AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.