FDA Adverse Event Malfunction Summary report: N

PULSE GEN MODEL 102R

MDR report key: 3800932 · Received May 8, 2014

Report

Report Number
1644487-2014-01227
Event Type
Malfunction
Date Received
May 8, 2014
Date of Event
July 11, 2006
Report Date
April 11, 2014
Manufacturer
CYBERONICS INC
Product Code
LYJ
PMA / PMN Number
P970003
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE GENERATOR WAS FOUND TO HAVE A LEAKY CAPACITOR, POTENTIALLY CONTRIBUTING TO A PREMATURE END OF SERVICE CONDITION.

Description of Event or Problem · 1

ANALYSIS OF A GENERATOR EXPLANTED FOR END OF SERVICE WAS COMPLETED ON (B)(4) 2014. THE SUPPLY CURRENT TESTS DID NOT MEET FUNCTIONAL SPECIFICATIONS. THESE MEASUREMENTS DEMONSTRATE AN INCREASED CURRENT CONSUMPTION FOR THE DEVICE, POTENTIALLY CONTRIBUTING TO A PREMATURE END OF LIFE CONDITION. THE INCREASED CURRENT CONSUMPTION WAS ISOLATED TO A LEAKY CAPACITOR (C4). A BATTERY LIFE ESTIMATION RESULTED IN 0.48 YEARS REMAINING BEFORE THE ERI FLAG WOULD BE SET. HOWEVER, AN INCOMPLETE PROGRAMMING HISTORY INDICATES THE ESTIMATION DOES NOT USE ALL THE DATA REQUIRED TO MAKE AN ACCURATE ESTIMATION. WITH THE CAPACITOR SUBSTITUTION FOR C4, THE PULSE GENERATOR MODULE PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE FOR THE PREMATURE END OF LIFE CONDITION WAS IDENTIFIED TO BE A LEAKY CAPACITOR, C4. THE CAUSE FOR THE C4 CAPACITORS INCREASE IN LEAKAGE COULD NOT BE DETERMINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
277531 PULSE GEN MODEL 102R GENERATOR LYJ CYBERONICS INC 102R 015709

Patients

Seq Age Sex Outcome Treatment
1 46 YR