FDA Adverse Event Malfunction Summary report: N

PK DISSECTING FORCEPS INSTRUMENT

MDR report key: 3800921 · Received May 8, 2014

Report

Report Number
2955842-2014-02871
Event Type
Malfunction
Date Received
May 8, 2014
Date of Event
April 16, 2014
Report Date
April 16, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K061260
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTUITIVE SURGICAL, INC. (ISI) HAS RECEIVED THE INSTRUMENT INVOLVED WITH THE COMPLAINT AND COMPLETED DEVICE EVALUATION. INVESTIGATION CONFIRMED THE CUSTOMER'S REPORTED ISSUE. THE INSTRUMENT PITCH CABLES WERE BROKEN AT THE PROXIMAL CLEVIS HUB. THE CLEVIS DOES NOT EXHIBIT ANY WEAR. THE FRAYED STRANDS STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGES WERE FOUND ON THE INSTRUMENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PK DISSECTING FORCEPS INSTRUMENT WIRE WAS BROKEN AND STICKING OUT AT THE END. NOTHING FELL INTO THE PATIENT AND THERE WAS NO REPORT OF PATIENT INJURY RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
277979 PK DISSECTING FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420227-03 N10140115 533

Patients

Seq Age Sex Outcome Treatment
1