FDA Adverse Event
Malfunction
Summary report: N
PK DISSECTING FORCEPS INSTRUMENT
MDR report key: 3800921
·
Received May 8, 2014
Report
- Report Number
- 2955842-2014-02871
- Event Type
- Malfunction
- Date Received
- May 8, 2014
- Date of Event
- April 16, 2014
- Report Date
- April 16, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K061260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
INTUITIVE SURGICAL, INC. (ISI) HAS RECEIVED THE INSTRUMENT INVOLVED WITH THE COMPLAINT AND COMPLETED DEVICE EVALUATION. INVESTIGATION CONFIRMED THE CUSTOMER'S REPORTED ISSUE. THE INSTRUMENT PITCH CABLES WERE BROKEN AT THE PROXIMAL CLEVIS HUB. THE CLEVIS DOES NOT EXHIBIT ANY WEAR. THE FRAYED STRANDS STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGES WERE FOUND ON THE INSTRUMENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PK DISSECTING FORCEPS INSTRUMENT WIRE WAS BROKEN AND STICKING OUT AT THE END. NOTHING FELL INTO THE PATIENT AND THERE WAS NO REPORT OF PATIENT INJURY RELATED TO THIS EVENT.