FDA Adverse Event Malfunction Summary report: N

EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM

MDR report key: 3800859 · Received May 8, 2014

Report

Report Number
2024168-2014-02942
Event Type
Malfunction
Date Received
May 8, 2014
Date of Event
April 17, 2014
Report Date
April 17, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
NTE
PMA / PMN Number
K090665
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A QUERY OF THE ELECTRONIC COMPLAINT HANDLING DATABASE REVEALED NO OTHER SIMILAR INCIDENTS REPORTED FROM THIS LOT. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE EMBOSHIELD NAV 6 OUTER PACKAGE WAS OPENED AND THE ENTIRE RETRIEVAL CATHETER POUCH WAS NOTED TO BE MISSING. THE DEVICE WAS DISCARDED AND NOT USED. THERE WAS NO PATIENT INVOLVEMENT AND ANOTHER EMBOSHIELD WAS USED IN REPLACEMENT. THERE WAS NO CLINICALLY SIGNIFICANT PROCEDURAL DELAY. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
277175 EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM EMBOLIC PROTECTION SYSTEM NTE AV-TEMECULA-CT 4012261

Patients

Seq Age Sex Outcome Treatment
1