FDA Adverse Event Malfunction Summary report: N

SAGITTAL SAW ATTACHMENT

MDR report key: 3800622 · Received May 8, 2014

Report

Report Number
8030965-2014-10228
Event Type
Malfunction
Date Received
May 8, 2014
Date of Event
March 1, 2014
Report Date
April 16, 2014
Manufacturer
SYNTHES OBERDORF
Product Code
HWE
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO IMMERSION DURING CLEANING. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SAGITTAL SAW ATTACHMENT DEVICE WOULD NOT WORK. THE EVENT WAS NOT REPORTED TO HAVE OCCURRED DURING SURGERY. THE REPORTER STATED THAT THE DEVICE WAS LEFT WITH A NOTE FOR REPAIR. IT WAS UNKNOWN TO THE REPORTER IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN, HOWEVER, THE REPORTER STATED THAT THE EVENT OCCURRED DURING THE MONTH OF (B)(6) 2014. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
279459 SAGITTAL SAW ATTACHMENT INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT HWE SYNTHES OBERDORF 9195

Patients

Seq Age Sex Outcome Treatment
1