FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3800611
·
Received February 11, 2014
Report
- Report Number
- 8020893-2014-00329
- Event Type
- Malfunction
- Date Received
- February 11, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 28, 2014
- Manufacturer
- COVIDIEN, FORMLERY NELLCO
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE GRAPHIC USER INTERFACE (GUI) CENTRAL PROCESSING UNIT (CPU) PRINTED CIRCUIT BOARD (PCB). THE CSE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED.
Description of Event or Problem · 1
COVIDIEN RECEIVED INFORMATION STATING THAT AN 840 VENTILATOR HAD A BLANK SCREEN. THE VENTILATOR WAS NOT IN USE ON A PT AT THE TIME THE MALFUNCTION OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 90106 | 840 VENTILATOR | CBK: VENTILATOR, CONTINUOUS, FACILITY | CBK | COVIDIEN, FORMLERY NELLCO | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |