FDA Adverse Event Malfunction Summary report: N

CONTOUR NEXT

MDR report key: 3800610 · Received May 8, 2014

Report

Report Number
1826988-2014-00159
Event Type
Malfunction
Date Received
May 8, 2014
Date of Event
April 1, 2014
Report Date
April 15, 2014
Manufacturer
BAYER HEALTHCARE LLC
Product Code
NBW
PMA / PMN Number
K110894
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER RETURNED A SECOND LOT# OF CONTOUR NEXT TEST STRIPS FOR EVALUATION, WHICH WAS NOT INITIALLY PROVIDED. LOT# 3FFEF02, EXPIRATION DATE 06/30/2015. MANUFACTURE DATE 06/01/2013. THE EVALUATION RESULTS FOR THIS LOT#: USING CUSTOMER'S CONTROL SOLUTION, THE READINGS WERE AN AVERAGE OF 3MG/DL HIGH OUT OF SPEC. THE RETENTION REAGENT GAVE SATISFACTORY PERFORMANCE.

Description of Event or Problem · 1

THE ADVOCATE STATED HER SON RECEIVED A BLOOD GLUCOSE READING OF 199MG/DL FROM THE CONTOUR NEXT LINK METER, RE-TESTED ON A DIFFERENT METER AND RECEIVED 54MG/DL.THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT.NO ADVERSE EVENT WAS ALLEGEDTHE TEST STRIPS WERE RETURNED FOR EVALUATION.REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
279386 CONTOUR NEXT BLOOD GLUCOSE TEST STRIPS NBW BAYER HEALTHCARE LLC 7311 3HFEC03

Patients

Seq Age Sex Outcome Treatment
1