FDA Adverse Event Malfunction Summary report: N

LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (SMALL)

MDR report key: 3800563 · Received February 11, 2014

Report

Report Number
2024601-2014-00033
Event Type
Malfunction
Date Received
February 11, 2014
Date of Event
April 19, 2013
Report Date
January 14, 2014
Manufacturer
ALLERGAN
Product Code
LTI
PMA / PMN Number
P000008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

TAPER II. ALLERGAN HAS REC'D THE PRODUCT HOWEVER THE DEVICE HAS NOT BEEN IDENTIFIED NOR HAS THE ANALYSIS HAS NOT BEEN COMPLETED AT THIS TIME. BASED UPON THE MODEL NUMBER AND IMPLANT DATE PROVIDED BY THE REPORTER THE CONNECTOR TYPE IS ASSUMED TO BE A TAPER II. VISUAL EXAMINATION MAY DETERMINE THE CONNECTOR TYPE ASSOCIATED WITH THIS REPORT. NO ADD'L INFO HAS BEEN REPORTED TO ALLERGAN REGARDING THE SERIAL NUMBER. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."

Description of Event or Problem · 1

EXPLANTING PHYSICIAN REPORTS "LEAK FROM MID PORTION OF THE TUBE. BAND TUBE FIRST LEAK, THEN SNAPPED WHEN PULLED." THE BAND WAS REPLACED WITH ANOTHER APS LAP-BAND SYSTEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
90011 LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (SMALL) LTI ALLERGAN NA NI

Patients

Seq Age Sex Outcome Treatment
1 39 YR