PRECISE BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-02854
- Event Type
- Malfunction
- Date Received
- May 8, 2014
- Date of Event
- March 18, 2014
- Report Date
- April 15, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHARMACIST
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION WAS UNABLE TO CONFIRM THE REPORTED COMPLAINT OF THE INSTRUMENT NOT FUNCTIONING. THE INSTRUMENT WAS PLACED ON AN IN-HOUSE SYSTEM, AND THE INSTRUMENT WAS SUCCESSFULLY RECOGNIZED. THE INSTRUMENT MOVED INTUITIVELY WITH FULL RANGE OF MOTION IN ALL DIRECTIONS, AND THE GRIPS OPENED AND CLOSED PROPERLY. FAILURE ANALYSIS INVESTIGATION FOUND THAT THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS EXHIBITING LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE NOT AXIALLY ALIGNED WITH THE TUBE. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE TUBE ABRASIONS WITH MATERIAL REMOVED, FOUND DURING FAILURE ANALYSIS INVESTIGATION, COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTIONS WERE TO RECUR.
IT WAS REPORTED THAT DURING A DA VINCI SURGICAL PROCEDURE, THE PRECISE BIPOLAR FORCEPS INSTRUMENT WAS RECOGNIZED BY THE SYSTEM, BUT DID NOT FUNCTION. THE ENERGY CABLE WAS EXCHANGED, BUT THE INSTRUMENT STILL DID NOT FUNCTION. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 277926 | PRECISE BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420110-06 | M10130711 476 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |