FDA Adverse Event Malfunction Summary report: N

PRECISE BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3800481 · Received May 8, 2014

Report

Report Number
2955842-2014-02854
Event Type
Malfunction
Date Received
May 8, 2014
Date of Event
March 18, 2014
Report Date
April 15, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION WAS UNABLE TO CONFIRM THE REPORTED COMPLAINT OF THE INSTRUMENT NOT FUNCTIONING. THE INSTRUMENT WAS PLACED ON AN IN-HOUSE SYSTEM, AND THE INSTRUMENT WAS SUCCESSFULLY RECOGNIZED. THE INSTRUMENT MOVED INTUITIVELY WITH FULL RANGE OF MOTION IN ALL DIRECTIONS, AND THE GRIPS OPENED AND CLOSED PROPERLY. FAILURE ANALYSIS INVESTIGATION FOUND THAT THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS EXHIBITING LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE NOT AXIALLY ALIGNED WITH THE TUBE. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE TUBE ABRASIONS WITH MATERIAL REMOVED, FOUND DURING FAILURE ANALYSIS INVESTIGATION, COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTIONS WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SURGICAL PROCEDURE, THE PRECISE BIPOLAR FORCEPS INSTRUMENT WAS RECOGNIZED BY THE SYSTEM, BUT DID NOT FUNCTION. THE ENERGY CABLE WAS EXCHANGED, BUT THE INSTRUMENT STILL DID NOT FUNCTION. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
277926 PRECISE BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420110-06 M10130711 476

Patients

Seq Age Sex Outcome Treatment
1