FDA Adverse Event
Malfunction
Summary report: N
8CM ANGLE ATTACHMENT
MDR report key: 3800457
·
Received May 8, 2014
Report
- Report Number
- 1045834-2014-12002
- Event Type
- Malfunction
- Date Received
- May 8, 2014
- Date of Event
- April 16, 2014
- Report Date
- April 16, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- PK011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. A FUNCTIONAL ASSESSMENT WAS PERFORMED AND THE DEVICE MET ALL MANUFACTURE'S SPECIFICATIONS. THE REPORTED CONDITION WAS NOT CONFIRMED. THEREFORE, AN ASSIGNABLE ROOT CAUSE WAS NOT DETERMINED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING TESTING, THE ATTACHMENT DEVICE WAS "GETTING HOT". IT WAS UNKNOWN IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 277839 | 8CM ANGLE ATTACHMENT | MOTOR, DRILL, ELECTRIC - ATTACHMENT | HBC | DEPUY SYNTHES POWER TOOLS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |